Programming Manual
4.7.7 Source - Pulse subsystem
[:SOURce]
:PULSE
:COUNt 1 to 2e8
:DCYCle 0 to 100% pulse width must be >= 0.001
:HOLD WIDTh | DCYCle
:PERiod 0.002 to 90,000
:WIDth 0.001 to < 90,000
The source pulse sub-system is used to program the pulsed output mode of operation of the DC
Supply.
PULSe:COUNt
This command sets the number of pulses that are output when a triggered output transient
occurs. The command accepts parameters in the range 1 through 2E8. If MAXimum is sent, the
output pulse repeats indefinitely.
Command Syntax [SOURce:]PULSe:COUNt<NRf+>
Parameters 1 to 2E8 | MINimum | MAXimum
*RST Value 1
Examples PULS:COUN 3
PULS:COUN MIN PULS:COUN MAX
Query Syntax [SOURce:]PULS:COUNt?
Returned Parameters <NR2>
Related Commands PULS:DCYC PULS:PER PULS:WIDT
PULSe:DCYCle
This command sets the duty cycle of the triggered output pulse. The duty cycle units are
specified in percent.
Command Syntax [SOURce:]PULSe:DCYCle<NRf+>
Parameters 0 to 100%|MINimum|MAXimum
*RST Value 50%
Examples PULS:DCYC 75
PULS:DCYC MAX
Query Syntax [SOURce:]PULSe:DCYCle?
Returned Parameters <NR2>
Related Commands PULS:COUN PULS:PER PULS:WIDT
PULSe:HOLD
This command specifies whether the pulse width or the duty cycle is to be held constant when
the pulse period changes. The following tables describe how the duty cycle, period, and width
are affected when one parameter is set in a single program message. if two or three parameters
are set in a single message, there sequence will define how they will be affected.
KDC Series DC Power Supplies 57