Programming Manual
Command Syntax CALibrate:SOURce: VOLTage:EXTernal:10V <NRf+>
Parameters <NRf> (a value between -127 and +128)
Examples CAL:VOLT:EXT:10V -2
Query Syntax CALibrate:VOLTage:EXTernal:10V?
Returned Parameters <NR1> (value range -127 to +128)
Related Commands CAL:CURR:EXT:10V
CALibrate[:SOURce]:VOLTage:EXTernal:RESistance <NRf+>
This command will set the calibration coefficient for the RPV resistive input.
Command Syntax CALibrate:SOURce:VOLTage:EXTernal:RESistance <NRf+>
Parameters <NRf> (a value between -127 and +128)
Examples CAL:VOLT:EXT:RES -2
Query Syntax CALibrate:VOLTage:EXTernal:RESistance?
Returned Parameters <NR1> (value range -127 to +128)
Related Commands CAL:CURR:EXT:RES
CALibrate[:SOURce]:VOLTage:INTernal <NRf+>
This command will set the calibration coefficient for the internal voltage range.
Command Syntax CALibrate:SOURce:VOLTage:INTernal <NRf+>
Parameters <NRf> (a value between -127 and +128)
Examples CAL:VOLT:INT -2
Query Syntax CALibrate:VOLTage:INTernal?
Returned Parameters <NR1> (value range -127 to +128)
Related Commands CAL:CURR:INT
CALibrate[:SOURce]:VOLTage:OFFSet <NRf+>
This command will set the calibration coefficient for the dc offset current.
Command Syntax CALibrate:SOURce:VOLTage:OFFSet <NRf+>
Parameters <NRf> (a value between -127 and +128)
Examples CAL:VOLT:OFFS -2
Query Syntax CALibrate:VOLTage:OFFSet?
Returned Parameters <NR1> (value range -127 to +128)
Related Commands CAL:VOLT:RAN
CALibrate[:SOURce]:VOLTage:RANGe <NRf+>
This command will set the calibration coefficient for the dc full-scale output voltage.
Command Syntax CALibrate:SOURce:VOLTage:RANGe <NRf+>
Parameters <NRf> (a value between -127 and +128)
Examples CAL:VOLT:RANG -2
Query Syntax CALibrate:VOLTage:RANGe?
Returned Parameters <NR1> (value range -127 to +128)
Related Commands CAL:VOLT:OFFS
KDC Series DC Power Supplies 31