Programming Manual
4.11.1 *CLS
This command clears the following registers (see chapter 6 for descriptions of all status
registers):
 Standard Event Status
 Status Byte
 Error Queue
Command Syntax *CLS
Parameters None
4.11.2 *ESE
This command programs the Standard Event Status Enable register bits. The programming
determines which events of the Standard Event Status Event register (see *ESR?) are allowed
to set the ESB (Event Summary Bit) of the Status Byte register. A "1" in the bit position enables
the corresponding event. All of the enabled events of the Standard Event Status Enable Register
are logically OR-ed to cause the Event Summary Bit (ESB) of the Status Byte Register to be set.
See Section 16 for descriptions of the Standard Event Status registers.
The query reads the Standard Event Status Enable register.
Table 4-3: Bit Configuration of Standard Event Status Enable Register
Bit Position 7 6 5 4 3 2 1 0
Bit Name PON not used CME EXE DDE QYE not
used
not used
Bit Weight 128 32 16 8 4
CME Command error DDE Device-dependent error
EXE Execution error QYE Query error
PON Power-on
Command Syntax *ESE <NRf>
Parameters 0 - 255
Power-On Value 128
Example *ESE 129
Query Syntax *ESE?
Returned Parameters <NR1>(Register value)
Related Commands *ESR? *STB?