Programming Manual
FIXed The current slew level set is unaffected by a triggered output transient.
STEP The current slew level set is programmed to the value set by
CURRent:SLEW:TRIGgered when a triggered transient occurs.
PULSe The current slew level set is changed to the value set by
CURRent:SLEW:TRIGgered for a duration determined by the pulse commands.
LIST The current slew level set is controlled by the current slew list when a triggered
transient occurs.
The current slew may be set between zero and the model's maximum current slew capability
(determined by model) which can be queried with the CURR:SLEW? MAX command.
Command Syntax [SOURce:]CURRent:SLEW:MODE <mode>
Parameters FIXed | STEP | PULSe | LIST
*RST Value FIX
Examples CURR:SLEW:MODE LIST CURR:MODE FIX
Query Syntax [SOURce:]CURRent:SLEW:MODE?
Returned Parameters <CRD>
Related Commands CURR:TRIG CURR
[SOURce:]CURRent:SLEW:TRIGgered
This command sets the current slew rate trigger level. When triggered by a step or pulse
transient, the current slew rate will be set to the programmed level. Note that the new slew rate
does not take effect until the unit is armed and a trigger event occurs. A parameter of MAXimum
will set the slew to its maximum possible rate.
The current slew may be set between zero and the model's maximum current slew capability
(determined by model) which can be queried with the CURR:SLEW? MAX command.
Command Syntax [SOURce:]CURRent:SLEW:TRIGgered <NRf>
Parameters 0 to MAX
*RST Value 0
Examples CURR:SLEW:TRIG
Query Syntax [SOURce:]CURRent:SLEW:TRIG?
Returned Parameters <NR2>
Related Commands CURR:SLEW:MODE CURR:SLEW
[SOURce:]CURRent:TRIGgered[:AMPLitude]
This command sets the current trigger level. When triggered by a step or pulse transient, the
current will be set to the programmed level. Note that the new current setting does not take
effect until the unit is armed and a trigger event occurs. The maximum current that can be
programmed is a function of the model and may be queried using the LIMIT subsystem. It is
also further determined by the user defined CURR:HIGH and CURR:LOW settings.
The current slew may be set between zero and the model's maximum current slew capability
(determined by model) which can be queried with the CURR:SLEW? MAX command.
Command Syntax [SOURce:]CURRent:TRIGgered[:AMPLitude <NRf>
Parameters 0 to MAX
*RST Value 0
Examples CURR:TRIG
Query Syntax [SOURce:]CURRent: TRIG?
Returned Parameters <NR2>
Related Commands CURR:MODE CURR
KDC Series DC Power Supplies 49