Programming Manual
Command Syntax [SOURce:]CURRent:HIGH <NRf>
Parameters 0 to LIM:CURR? value.
Unit A (amperes)
*RST Max. Current
Examples CURR:HIGH 5
Query Syntax CURRent:HIGH?
Returned Parameters <NR2>
Related Commands CURR CURR:LOW
[SOURce:]CURRent[:LEVel][:IMMediate]:LOW
This command sets the minimum programmed current that will be accepted by the power
supply. If a value below this limit value is programmed, a "-222 Data out of Range" error will be
generated.
The current limit may be set between zero and the model's capability (configuration limit) which
can be queried with the LIM:CURR? command. The low limit cannot be more than the high limit
however. The *RST state for low current limit is zero.
Command Syntax [SOURce:]CURRent:LOW <NRf>
Parameters 0 to LIM:CURR? value.
Unit A (amperes)
*RST 0
Examples CURR:LOW 5
Query Syntax CURRent:LOW?
Returned Parameters <NR2>
Related Commands CURR CURR:HIGH
[SOURce:]CURRent:MODE
This command sets the current programming mode. Available modes are:
FIXed The current level set is unaffected by a triggered output transient.
STEP The current level set is programmed to the value set by CURRent:TRIGgered
when a triggered transient occurs.
PULSe The current level set is changed to the value set by CURRent:TRIGgered for a
duration determined by the pulse commands.
LIST The current level set is controlled by the current list when a triggered transient
occurs.
The current limit may be set between zero and the model's capability (configuration limit) which
can be queried with the LIM:CURR? command. The low limit cannot be more than the high limit
however. The *RST state for low current limit is zero.
Command Syntax [SOURce:]CURRent:MODE <mode>
Parameters FIXed | STEP | PULSe | LIST
*RST Value FIX
Examples CURR:MODE LIST CURR:MODE FIX
Query Syntax [SOURce:]CURRent:MODE?
Returned Parameters <CRD>
Related Commands CURR:TRG CURR
[SOURce:]CURRent:PROTection:DELay
This command sets the programmable current limit delay when in CV mode of operation. The
delay determines for how long the shutdown of the output will be delayed. The programmable
range is from 0.1 (100 msec) to 5 secs.
KDC Series DC Power Supplies 47